RX Measurements for CA 3.4.
The following test procedures are different between the MT8820C and MT8821C.
This chapter explains each test procedure for the MT8820C and MT8821C.
Reference sensitivity level for CA (intra-band contiguous DL CA and UL CA) (7.3A.1) 3.4.1.
3.4.1.1. MT8820C
1. [PCC/SCC] Execute TESTPRM RX_SENS to set Test Parameter to RX - Ref. Sens./Freq. Error.
2. [PCC] Execute ULRB_START 0 to set UL RMC Starting RB to 0.
3. [PCC] Execute TPUT SAMPLE 10000 to set the Throughput measurement sample count to 10000.
4. [PCC] Execute TPUT_EARLY ON to set Early Decision to On.
5. [PCC] Execute TPUT_EARLY_TARCC PCC_SCC to set Target CC to PCC+SCC.
6. [PCC] Execute SWP to measure the Throughput.
7. [PCC] Execute TPUTPASS? to check that the Throughput measurement Pass/Fail judgment is Pass.
3.4.1.2. MT8821C
1. Execute TESTPRM RX_SENS to set Test Parameter to RX Ref. Sens./Freq. Error.
2. Execute MOD_MEAS OFF to set Modulation Analysis Measurement to OFF.
3. Execute ULRB_START 0 to set Common Parameter UL RMC - Starting RB to 0.
4. Execute TPUT SAMPLE 10000 to set Rx Measurement Parameter Throughput Number of Sample to
10000.
5. Execute TPUT_EARLY ON to set Rx Measurement Parameter Throughput Early Decision to On.
6. Execute TPUT_EARLY_TARCC PCC_SCC to set Rx Measurement Parameter Throughput Early Decision -
Target CC to PCC+SCC.
7. Execute SWP to measure the Throughput.
8. Execute TPUT? PER to read the Throughput measurement result (%)
9. Execute TPUTPASS? to check that the Throughput measurement Pass/Fail judgment is Pass.
NOTE: Since PCC/SCC1 UL RMC - Number of RB depends on the Operation Band, set the value in
TS36.5211 Table 7.3A.1.4.11 and 7.3A.1.31.
Reference sensitivity level for CA (intra-band contiguous DL CA without UL CA) 3.4.2.
(7.3A.2)
Refer to Chapter 3.4.1.
NOTE: Since PCC/SCC1 UL RMC Number of RB depends on the Operation Band, set the value in
TS36.5211 Table 7.3A.2.4.11.