Maximum input level for CA (intra-band contiguous DL CA without UL CA) (7.4A.2) 3.4.6.
Refer to Chapter 3.4.5 except for step 2.
NOTE 1: Since UL RMC Number of RB/Starting RB depends on the CA Configuration, set the Number of
RB/Starting RB value for each CC according to TS36.5211 Table 7.4A.2.4.11 and Table
7.3A.1.31.
NOTE 2: The output power for each CC depends on the Transmission Bandwidth and the Carrier
Frequency f as described in TS36.5211.
Power in largest transmission bandwidth CC
f ≤ 3.0 GHz: 25.7 dBm
3.0 GHz < f ≤ 4.2 GHz: 26.0 dBm
Power in each other CC
f ≤ 3.0 GHz: 25.7 + 10Log(N
RB,c
/N
RBlargestBW
) dBm
3.0 GHz < f ≤ 4.2 GHz: 26.0 + 10Log(N
RB,c
/N
RBlargestBW
) dBm
Maximum input level for CA (inter-band DL CA without UL CA) (7.4A.3) 3.4.7.
This chapter describes the measurement examples for intra-band measurement.
Example: PCC DL Channel = 300 (Band1), SCC DL Channel is 6075
PCC N
RB
= 100, SCC N
RB
= 50, N
RB_alloc
= 50,
PCC and SCC DL allocations (L
CRB
@RB
start
) are P_100@0 and S_50@0, respectively.
PCC and SCC UL allocations (L
CRB
@RB
start
) are P_100@0 and S_0@0, respectively.
This measurement can be performed using the same procedure as in Chapter 3.4.6, by substituting the following
steps.
1. Execute TESTPRM RX_MAX to set Test Parameter to RX Max. Input Level.
2. Execute ULRMC_RB 50 to set Common Parameter - UL RMC Number of RB to 50.
3. Execute TPUT_SAMPLE 10000 to set Rx Measurement Parameter Throughput Number of Sample to
10000.
4. Execute TPUT_EARLY ON to set Rx Measurement Parameter Throughput Early Decision to On.
5. Execute TPUT_EARLY_TARCC SCC to set Rx Measurement Parameter Throughput Early Decision -
Target CC to SCC.
6. Execute SWP to measure the Throughput.
7. Execute TPUTPASS? to check that the Throughput measurement Pass/Fail judgment is Pass.
8. Execute DLCHAN 6075,300 to set UL and DL Channel for PCC to 24075 and 6075, and switch the PCC and
SCC Channel.
9. Execute steps 6 to 7.
Maximum input level for CA (intra-band non-contiguous DL CA without UL CA) 3.4.8.
(7.4A.4)
This measurement can be performed using the same procedure as in Chapter 3.4.6, by substituting the following
steps.
10. Execute TPUT_EARLY_TARCC SCC to set Rx Measurement Parameter Throughput Early Decision
Target CC to SCC.
11. Execute SWP to measure the Throughput.
12. Execute TPUT? PER to read the Throughput measurement result (%).
13. Execute TPUTPASS? to check that the Throughput measurement Pass/Fail judgment is Pass.
14. Execute DLCHAN 0,300 to set UL and DL Channel for PCC to 18000 and 0, and switch the PCC and SCC
Channel.
15. Execute steps 11 to 13.