96
In-band emissions for non allocated RB - PUCCH (6.5.2.3) 3.1.25.
1. Execute MOD_AVG 20 to set the average count of Modulation Analysis to 20 times.
2. Execute TESTPRM TX_PUCCH_0DBM to set Test Parameter to TX2 - PUCCH IBE @ 0 dBm.
3. Execute SWP to measure Modulation Analysis.
4. Execute INBANDE_GEN? MAX to read the In-band Emissions (General) measurement result.
5. Execute INBANDE_IMG? MAX to read the In-band Emissions (IQ Image) measurement result.
6. Execute INBANDE_LEAK? MAX to read the In-band Emissions (Carrier Leakage) measurement result.
7. Execute INBANDEPASS? to check that the In-band Emissions Pass/Fail judgment is Pass.
8. Execute TESTPRM TX_PUCCH_M30DBM to set Test Parameter to TX2 - PUCCH IBE @ -30 dBm.
9. Execute steps 3 to 7.
10. Execute TESTPRM TX_PUCCH_M40DBM to set Test Parameter to TX2 - PUCCH EVM/IBE @ -40 dBm.
11. Execute steps 3 to 7.
NOTE 1: The UL RB Position for PartialRB allocation is Min (#0) or Max (#max).
Figure 3.1.25-1 Example of Measurement Result when Test Parameter is TX2 - PUCCH IBE @ 0 dBm (MT8820C)