Anritsu MT8820C - Carrier Leakage (6.5.2.2); In-Band Emissions for Non Allocated RB - PUSCH (6.5.2.3)

Anritsu MT8820C
268 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
94
Carrier leakage (6.5.2.2) 3.1.23.
This chapter describes a UL measurement example where (Modulation, RB) is (QPSK, PartialRB).
1. Execute MOD_AVG 20 to set the average count of Modulation Analysis to 20 times.
2. Execute TESTPRM TX_0DBM to set Test Parameter to TX1 - IBE/LEAK @ 0 dBm.
3. Execute ULRB_POS MIN to set UL RB Position to Min (#0).
4. Execute SWP to measure Modulation Analysis.
5. Execute CARRLEAK? MAX to read the Carrier Leakage measurement result.
6. Execute CARRLEAKPASS? to check that the Carrier Leakage Pass/Fail judgment is Pass.
7. Execute ULRB_POS MAX to set UL RB Position to Max (#max)
8. Execute steps 4 to 6.
9. Execute TESTPRM TX_M30DBM to set Test Parameter to TX1 - IBE/LEAK @ -30 dB.
10. Execute steps 3 to 8.
11. Execute TESTPRM TX_M40DBM_Q_P to set Test Parameter to TX1 - EVM/IBE/LEAK @ -40 dBm
(QPSK/PartialRB).
12. Execute steps 3 to 8.
NOTE 1: The UL RB Position of PartialRB allocation is Min (#0) or Max (#max).
In-band emissions for non allocated RB - PUSCH (6.5.2.3) 3.1.24.
This chapter describes a UL measurement example where (Modulation, RB) is (QPSK, PartialRB).
1. Execute MOD_AVG 20 to set the average count of Modulation Analysis to 20 times.
2. Execute TESTPRM TX_0DBM to set Test Parameter to TX1 - IBE/LEAK @ 0 dBm.
3. Execute ULRB_POS MIN to set UL RB Position to Min (#0).
4. Execute SWP to measure Modulation Analysis.
5. Execute INBANDE_GEN? MAX to read the In-band Emissions (General) measurement result.
6. Execute INBANDE_IMG? MAX to read the In-band Emissions (IQ Image) measurement result.
7. Execute INBANDE_LEAK? MAX to read the In-band Emissions (Carrier Leakage) measurement result.
8. Execute INBANDEPASS? to check that the In-band Emissions Pass/Fail judgment is Pass.
9. Execute ULRB_POS MAX to set UL RB Position to Max (#max).
10. Execute steps 4 to 8.
11. Execute TESTPRM TX_M30DBM to set Test Parameter to TX1 - IBE/LEAK @ -30 dBm.
12. Execute steps 3 to 10.
13. Execute TESTPRM TX_M40DBM_Q_P to set Test Parameter to TX1 - EVM/IBE/LEAK @ -40 dBm
(QPSK/PartialRB).
14. Execute steps 3 to 10.
NOTE 1: The UL RB Position for PartialRB allocation is Min (#0) or Max (#max).