72
Location Registration 2.4.3.
This performs UE location registration after setting the initial conditions (2.4.3).
2.4.3.1. MT8821C
1. Connect the UE and MT8821C.
2. Execute CALLPROC ON to set Common Parameter – Call processing to ON.
3. Execute CALLSO to clear the call processing status.
4. Execute CALLSTAT? to confirm the call processing status is 1 (= Idle).
5. Turn on the UE power.
6. Execute CALLSTAT? to confirm the call processing status is 2 (= Idle (Regist)).
Repeat step 6 when the checked status is not 2 (= Idle (Regist)).
Test Mode Connection and Disconnection
2.4.4.
Refer to Chapter 2.1.4.