242
Measurement Report 6.3.
This chapter describes the Measurement Report.
This function can verify the Inter-RAT measurement function and receiver characteristics for neighboring cells, such
as E-UTRA inter-frequency/intra-frequency, Inter-RAT UTRA FDD/TDD, GSM, CDMA2000 in RRC_CONNECTED state.
Initial Condition Setting 6.3.1.
In this and following chapters, the initial settings are 480 ms for Measurement Report - Interval, and Periodical for
Measurement Report - Trigger Type. Changes can be made if necessary.
1. Execute MEASREP_INTVAL 480 to send UE Report every 480 ms.
2. Execute MEASREP_TRG PERIODICAL to set Measurement Report - Trigger Type to Periodical.
Measurement Report Procedure 6.3.2.
6.3.2.1. Measurement Report for LTE
This chapter describes the UE Report for the Serving Cell (LTE).
1. Execute CALLRFR to initialize the UE Report value.
2. Execute NEIGHCELLMEAS OFF to set Neighbour Cell Measurement to OFF.
3. Connect in Test Mode. (2.3)
4. Execute MEASREP ON to request UE Report.
5. Execute RSRP? FLAG and check if the Response is 1 to receive the UE Report.
6. Execute RSRP? to read the RSRP value.
7. To read again, return to Step 4 and continue.
For a one-time UE Report, the changes to the procedure are shown below.
These changes can be applied to the following procedures.
5. Execute MEASREP OFF.
6. Execute MEASREP_ONCE to request UE report only once.
6.3.2.2. Measurement Report for Intra-Frequency
This chapter describes the UE Report for LTE and Intra-Frequency. In the example, the Cell ID for Intra-Frequency is
set to 100 and TAC is set to 000A. A UE Report for the Neighbour Cell requires signal input to the UE. Signal Input is
also required for the following procedures.
1. [LTE] Execute CALLRFR to initialize the UE Report value.
2. [LTE] Execute NCAINTRAFREQ 1,100 to set leftmost Intra Frequency Cell - Cell ID to 100.
3. [LTE] Execute NEIGHCELLMEAS INTRAFREQ to set Neighbour Cell Measurement to Intra Frequency.
4. [IntraFreq] Execute CELLID 100 to set Cell ID to 100.
5. [IntraFreq] Execute TAC 000A to set TAC to 000A.
6. [LTE] Connect in the Test Mode. (2.3)
7. [LTE] Execute MEASREP ON to request UE Report.
8. [LTE] Execute MREP_LTE? FLAG and check if the Response is 1 to receive the UE Report for Intra-Frequency.
9. [LTE] Execute MREP_LTE? to read the Cell ID, RSRP, and RSRQ values.
10. To read again, return to Step 8 and continue.