3.2.3. Spurious emissions (7.9)
Perform Rx spurious emission tests using an external spectrum analyzer.
1. Connect the MT8821C, spectrum analyzer and UE.
2. Execute CALLDROP OFF to set Call Processing Parameter - Call Drop function to OFF.
3. Execute ULRMC_RB 0 to set Common Parameter - UL RMC - Number of RB to 0.
4. Execute DLRMC_RB 0 to set Common Parameter - DL RMC - Number of RB to 0.
5. Measure the Rx spurious emissions using the spectrum analyzer.
6. Check that maximum level at each frequency bandwidth is lower than the standardized value.
NOTE 1: Refer to 3GPP TS36.508 Annex A, Figure A.8 for the connection between the MT8821C,
spectrum analyzer and UE.