[(QPSK, FullRB) measurements]
20. Execute ULRMC_RB 25 to set UL RB number to 25.
21. Execute TP_MPR1_UL 33.7 to set TX Power measurement Pass/Fail upper limit to 33.7 dBm.
22. Execute TP_MPR1_LL 26.3 to set TX Power measurement Pass/Fail lower limit to 26.3 dBm.
23. Execute steps 14 to 17.
[(16QAM, PartialRB) measurements]
24. Execute ULRMC_MOD 16QAM to set UL RMC modulation method to 16QAM.
25. Execute ULRMC_RB 8 to set UL RB number to 8.
26. Execute ULRB_POS MIN to set UL RB Position to Min (#0).
27. Execute TP_MPR1_UL 33.7 to set TX Power measurement Pass/Fail upper limit to 33.7 dBm.
28. Execute TP_MPR1_LL 26.3 to set TX Power measurement Pass/Fail lower limit to 26.3 dBm.
29. Execute steps 14 to 17.
30. Execute ULRB_POS MAX to set UL RB Position to Max (#max).
31. Execute steps 14 to 17.
NOTE 1: The UL RB Position for PartialRB allocation is divided as follows:
When Test Frequency is Low range, Max (#max)
When Test Frequency is Mid range, Min (#0) and Max (#max)
When Test Frequency is High range, Min (#0)
NOTE 2: There is no need to set separately because the Pass/Fail evaluation value for Spectrum Emission
Mask measurement changes in accordance with the additionalSpectrumEmission setting.
NOTE 3: The usual Pass/Fail evaluation value is set to the Band 1 default value described in TS36.521-1.
Since the evaluation value differs according to the Band, set the evaluation value described in
TS36.521-1 Table 6.2.4_1.5-1 at:
TP_MPR1_LL
TP_MPR1_UL
For the Pass/Fail evaluation values, refer to Chapter 3.7.4 Test Parameter Limit in the
operation manual.
Additional Spectrum Emission Mask (6.6.2.2) 3.1.35.
This measurement can be performed using the same procedure as in Chapter 3.1.33.