Adjacent Channel Leakage Power Ratio for HPUE (6.6.2.3_1) 3.1.31.
This measurement can be performed using the same procedure as in Chapter 3.1.30, except the Pass/Fail
evaluation limits value setting.
[Pass/Fail evaluation limits value setting]
1. Execute TP_ACLR_E -36.2 to set E-UTRA Pass/Fail limit to –36.2 dB.
2. Execute TP_ACLR_U1 0 UTRA
ACLR1
to set Pass/Fail limit to 0 dB.
3. Execute TP_ACLR_U2 0 UTRA
ACLR1
to set Pass/Fail limit to 0 dB.
Adjacent Channel Leakage Power Ratio for Multi-Cluster PUSCH (6.6.2.3_2) 3.1.32.
This chapter describes measurement examples for following conditions.
First example: Channel Bandwidth = 20 MHz, UL Modulation is 16QAM,
UL Number of RB and Starting RB of Cluster1 is 4,0 respectively and
UL Number of RB and Starting RB of Cluster2 is 4,96 respectively.
Second example: Channel Bandwidth = 20 MHz, UL Modulation is 16QAM,
UL Number of RB and Starting RB of Cluster1 is 92,0 respectively and
UL Number of RB and Starting RB of Cluster2 is 4,96 respectively.
[Pass/Fail evaluation limits value setting]
1. Execute ACLR_AVG 20 to set the average count of Adjacent Channel Power to 20 times.
2. Execute TP_ACLR_E -36.2 to set E-UTRA Pass/Fail limit value to –36.2 dB.
3. Execute TP_ACLR_U1 -32.2 UTRA
ACLR1
to set Pass/Fail limit value to –32.2 dB.
4. Execute TP_ACLR_U2 -35.2 UTRA
ACLR1
to set Pass/Fail limit value to –35.2 dB.
[(16QAM, PartialRB) measurements for first example]
5. Execute TESTPRM TX_MAXPWR_16_P to set Test Parameter toTX1 - Max. Power (16QAM/PartialRB)).
6. Execute CHCONFIG PUSCH_MULTI to set Common Parameter - RMC Configuration to PUSCH(Multi
Cluster).
7. Execute ULRB_MULTI 4,0,4,96 to set Common Parameter - UL RMC - 1
st
PUSCH Number of RB to 4, 1
st
PUSCH Starting RB to 0, 2
nd
PUSCH Number of RB to 4, and 2
nd
PUSCH Starting RB to 96.
8. Execute SWP to measure the Adjacent Channel Power.
9. Execute MODPWRPASS? to check that the ACLR Pass/Fail judgment is Pass.
[(16QAM, PartialRB) measurements for second example]
1. Execute ULRB_MULTI 92,0,4,96 to set Common Parameter - UL RMC - 1
st
PUSCH Number of RB to 92, 1
st
PUSCH Starting RB to 0, 2
nd
PUSCH Number of RB to 4, and 2
nd
PUSCH Starting RB to 96.
2. Execute steps 8 and 9.