3.3.2.18. Adjacent Channel Leakage power Ratio for CA (intra-band contiguous DL CA and UL CA)
(6.6.2.3A.1)
3.3.2.18.1. MT8820C
This measurement item is not supported by the MT8820C.
Refer to Table 3.31.
3.3.2.18.2. MT8821C
This subsection describes an example of intraband measurement.
[Measurements]
1. Execute ACLR_AVG 20 to set the average count for Adjacent Channel Leakage Ratio to 20 times.
2. Execute TESTPRM TX_MAXPWR_Q_F to set Test Parameter to TX1 Max. Power (QPSK/FullRB).
3. Execute SWP to measure the Adjacent Channel Power.
4. Execute TTL_MODPWR? to read the ACLR measurement result.
5. Execute MODPWRPASS? to check that the ACLR Pass/Fail judgment is Pass.