6.3.2.8. Measurement Report for CA
This chapter describes the UE Report for CA. In the example, the initial condition is set (2.2.3) and the UE Report
is requested when SCC Activation is Off and On.
1. Execute CALLRFR to initialize the UE Report value.
2. Execute MEASCYCLE_SCC1 SF1280 to set SCell Measurement Cycle of SCC-1 to sf1280.
3. Execute ACT_SCC1 OFF to set SCC-1 Activation to OFF.
4. Connect in the Test Mode. (2.2.5)
5. Execute MEASREP ON to request UE Report.
6. Execute RSRP_SCC1? FLAG and check if the Response is 1 to receive the UE Report for SCC-1.
7. Execute RSRP_SCC1? to read the SCC-1 RSRP value.
8. To read again, return to Step 7 and continue.
9. Execute MEASREP OFF to switch OFF transition for the UE Report.
10. Execute CALLRFR to initialize the UE Report value.
11. Execute ACT_SCC1 ON to set SCC-1 Activation to ON.
12. Execute MEASREP ON to request the UE Report.
13. Execute RSRP_SCC1? FLAG and check if the Response is 1 to receive the UE Report for SCC-1.
14. Execute RSRP_SCC1? to read the RSRP value for SCC-1.
15. To read again, return to Step 14 and continue.