6.
Close the access door and ensure it is completely closed.
Note: The QuantStudio
Real-Time PCR Software displays an error message if
the door is not completely closed and latched, or if the sample block is not seated
correctly.
7.
Plug in and power on the QuantStudio
6 or 7 Flex Real-Time PCR System.
8.
In the QuantStudio
Real-Time PCR Software Home tab, click Instrument
Console.
9.
In the Instrument Console, select your instrument from the list of instruments,
then review the Heated Cover Firmware Version eld in the Instrument
Properties pane.
The installation is successful if the system powers on and if the Heated Cover
Firmware Version eld displays a version number.
10.
Check the calibration status of the system (see “View the calibration status of an
instrument“ on page 29).
If the calibrations for the new sample block/heated cover have expired, perform
the following calibrations in the specied order:
a.
ROI calibration
b.
Background calibration
c.
Uniformity calibration
d.
Dye calibration
e.
Normalization calibration (TaqMan
®
Array Card block only)
IMPORTANT! If the calibrations for the new sample block/heated cov
er are
valid, you do not need to recalibrate the QuantStudio
6 or 7 Flex Real-Time PCR
System.
Change the plate adapter
Change the plate adapter in the ev
ent of a hardware failure or if you want to change
the consumable format of the QuantStudio
6 or 7 Flex System.
WARNING! PHYSICAL INJUR
Y HAZARD. During instrument operation, the
plate adapter can reach 100. Allow the plate adapter to cool to room
temperature before removing.
You can change the plate adapter:
To change the consumable format of the QuantStudio
6 or 7 Flex System.
In the event of a hardware failure
When to change
the plat
e adapter
Appendix B Maint
enance
Change the plate adapter
B
QuantStudio
6 and 7 Flex Real-Time PCR Systems v1.6 Maintenance and Administration Guide
111

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Applied Biosystems QuantStudio 6 Flex and is the answer not in the manual?

Applied Biosystems QuantStudio 6 Flex Specifications

General IconGeneral
BrandApplied Biosystems
ModelQuantStudio 6 Flex
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals