Calibration consumable preparation
Create a background plate or array card ................................ 151
Create a custom dye plate for calibration ............................... 153
Create a background plate or array card
Whenever possible, use a Background Plate or the TaqMan
®
Array Background Buer
that is included with the spectral calibration kit. The plates/array cards supplied in the
kit contain a buer that accurately simulates the reagents used for PCR, and,
therefore, produces high-quality calibration data. If a background plate or array card
from a spectral calibration kit is not available, you can create one as described below.
96/384-Well plate sample block
MicroAmp
Optical 96/384-Well Reaction Plate
Safety glasses
Optical Adhesive Cover or Optical Flat Caps
Pipeor, 200-µL (with pipee tips)
Powder-free gloves
Deionized water
TaqMan
®
Array Card block
(QuantStudio
7 Flex System only)
TaqMan
®
Array Cards
TaqMan
®
Array Card Sealer
Centrifuge with array card buckets and array card carrier clips
Permanent marker or pen
Pipeor, 200-µL (with pipee tips)
Powder-free gloves
Safety glasses
Deionized water
F
Required
materials
QuantStudio
6 and 7 Flex Real-Time PCR Systems v1.6 Maintenance and Administration Guide
151

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Applied Biosystems QuantStudio 6 Flex and is the answer not in the manual?

Applied Biosystems QuantStudio 6 Flex Specifications

General IconGeneral
BrandApplied Biosystems
ModelQuantStudio 6 Flex
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals