Store calibration plates or array cards in a dark place until you are ready to use
them. The uorescent dy
es in the wells of calibration consumables are
photosensitive. Prolonged exposure to light can diminish the uorescence of the
dyes.
Do not allow the booms of the plates or array cards to become dirty. Fluids and
other contaminants that adhere to the booms of the consumables can
contaminate the sample block and cause an abnormally high background signal.
Conrm that your centrifuge is clean. Before centrifugation, wipe down the
bucket(s) using a tissue.
Vortex and centrifuge all calibration plates to ensure complete mixing and that all
reagents are contained at the boom of the wells. The calibration plates must be
well mixed and centrifuged before use.
Preparing array cards for instrument calibration
IMPORTANT! P
erform the following procedure only if you are verifying the
performance of a QuantStudio
7 Flex System with an array card sample block. The
QuantStudio
6 Flex System does not support the array card sample block.
QuantStudio
7 Flex System Array Card Spectral Calibration Dye Kit:
Array Cards, empty
Array Card Spectral Calibration Dye Kit, including: FAM
dye mix, VIC
dye mix, ROX
dye mix, ROI dye mix, Background Buer, FAM
/ROX
dye
mix, and VIC
/ROX
dye mix
TaqMan
®
Array Card Sealer
Centrifuge with array card buckets and array card carrier clips
Permanent marker or pen
Pipeor, 200-µL (with pipee tips)
Powder-free gloves
Safety glasses
IMPORTANT! W
ear powder-free gloves while creating the calibration array cards.
This procedure explains how to create all of the array cards required to calibrate the
QuantStudio
7 Flex System, but not all of them are required for a monthly
maintenance. Before preparing array cards for calibration, determine which
calibrations are required (see “Calibration and maintenance schedule“ on page 27).
Not
e: For more information on the array card loading procedure, contact Technical
Support.
1.
Remove the tubes of calibration solutions from the freezer, allow them to thaw,
then vortex the tubes to mix the contents well.
2.
Remove the array cards from their box and place them on a clean, dry surface.
Required
mat
erials
Fill the calibration
T
aqMan
®
Array
Cards
Chapter 2 Calibr
ate the instrument
Preparing array cards for instrument calibration
2
QuantStudio
6 and 7 Flex Real-Time PCR Systems v1.6 Maintenance and Administration Guide
37

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Applied Biosystems QuantStudio 6 Flex and is the answer not in the manual?

Applied Biosystems QuantStudio 6 Flex Specifications

General IconGeneral
BrandApplied Biosystems
ModelQuantStudio 6 Flex
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals