b.
Enter any comments you have in the Comments eld, click Finish, then
click Yes when prompted to save the calibration results.
Failed—Create another custom dye plate using the next dye concentration
greater than the concentration determined in “Determine optimum dye
concentration“ on page 153, then perform the calibration again.
Appendix F Calibr
ation consumable preparation
Create a custom dye plate for calibration
F
QuantStudio
6 and 7 Flex Real-Time PCR Systems v1.6 Maintenance and Administration Guide
157