4.
Click Next, then prepare for the calibration as instructed.
5.
From the boom of the Setup tab, enter the reagent information for the ROI plate
or array card that y
ou are using.
6.
Load the calibration plate or array card into the instrument:
a.
From the instrument touchscreen, tap to eject the instrument tray.
b.
Load the plate or array card into the plate holder so that:
Well A1 of the plate or array card is in the top-left corner of the plate
adapter.
The barcode faces the front of the instrument.
IMPORTANT! Plates and array cards should be loaded and unloaded by
operators who hav
e been warned of the moving parts hazard and have been
adequately trained.
c.
From the instrument touchscreen, tap to close the instrument tray.
7.
After loading the plate or array card, start the calibration:
a.
At the boom of the Setup tab, select Check the box when the ROI
calibration plate has been loaded, then click Next.
b.
In the Run screen, click Start run.
IMPORTANT! Do not aempt to open the access door during the run. The door
is locked while the instrument is in operation.
Chapter 2 Calibrate the instrument
ROI calibration
2
44
QuantStudio
6 and 7 Flex Real-Time PCR Systems v1.6 Maintenance and Administration Guide

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Applied Biosystems QuantStudio 6 Flex and is the answer not in the manual?

Applied Biosystems QuantStudio 6 Flex Specifications

General IconGeneral
BrandApplied Biosystems
ModelQuantStudio 6 Flex
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals