4-2
AX-4030
AUTION MAX AX-4030 OPERATING MANUAL
Frequency of Maintenance4.1
The following table lists the maintenance tasks required at directed intervals.
Wear protective gloves to prevent exposure to pathogenic microbes when
performing maintenance tasks marked with a checkmark ( ).
Discard liquid waste, used parts, and cleaning tools in accordance with
local regulations for biohazardous waste.
Alcohol is sometimes used to clean the instrument. Alcohol is readily
combustible, therefore handle it carefully and keep away from flames,
electrical sparks and sources of heat. Also, ventilate the room sufficiently
during use.
Maintenance task Cycle See page(s)
Waste box cleaning Every day or before reaching 400
measurements
4-3
Liquid waste disposal (drain bottle) Every day or before reaching 600
measurements
4-5
Feeder cleaning Every three days
4-6
Test strip stopper cleaning Every three days
4-8
Introduction tray washout Every three days 4-12
Washing solution replacement Every 600 measurements 4-15
Thermal printer paper replacement If red lines appear at both edges of the paper 4-17
Air filter washout Every month 4-18
Transport tray washout Every week 4-19
S.G. cell washout Every week
4-24
Washing bath washout Every month 4-26
Washing solution filter replacement Every month
4-29
Drain pinch valve tube replacement If drainage is leaking from the pinch valve
tube (e.g. “T280”)
4-30
White plate replacement If warning “W007” occurs
4-32