4.4.2 Washing the Transport Tray
Particles from the test strips can accumulate and sample residue can crystallize on the transport
tray as more urine tests are run. This may hinder smooth transportation of the test strips. Wash and
sterilize the tray once a week to keep it clean. Introduction tray cleaning can also be done in the
course of the following procedure.
Prepare: Alcohol, cloth, and protective gloves
Wear protective gloves to prevent exposure to pathogenic microbes.
Discard used cleaning tools and protective gloves in accordance with local
regulations for biohazardous waste.
IMPORTANT: Observe the instructions in this section when washing the transport tray.
Improper procedures may misalign the tray from the designated position. This may
hinder smooth transportation of the test strips, consequently producing incorrect
results.
NOTE: Do not use thinner or other organic solvents to wash the transport and introduction trays.
Do not subject the trays to ultrasonic cleaning. Inappropriate cleaning may deform or discolor
the trays, possibly making it unusable.
1 Move the nozzle.
Move the nozzle before performing the cleaning procedures.
1 On the standby screen, press Menu to go to the
[Main menu] screen.
2 Press 5 to go to the [Maintenance] screen.
3 Press 2 to go to the [Clean washing bath and tray]
screen.
Menu Meas No. Strip S.G. Cal
Main menu
(0000) 2006-01-09 11:00
1 Parameter settings
2 Results
3 Initial settings
4 Print
5 Maintenance
Go back
Maintenance
(5000) 2006-01-09 11:00
1 S.G. cell washout
2 Clean washing bath and tray
3 Check measurement
4 Maintenance information