4-20
AX-4030
AUTION MAX AX-4030 OPERATING MANUAL
D
The nozzle will move to the designated position.
2 Turn off the power.
1 Turn off the power when prompted.
REFERENCE: Trouble “T290” occurs if you open
the cover without turning off the power.
3 Wash and sterilize the introduction tray.
1 Take the introduction tray out of the instrument and
sterilize it.
See steps 2 to 5 in “4.2.5 Washing the
Introduction Tray” on page 4-12.
NOTE: The introduction and transport trays are
similar in shape. It is recommended to wash the
introduction tray first and keep it in another place.
When washing the two trays at the same time,
handle the trays with care not to confuse them.
Clean washing bath and tray
(5200) 2006-01-09 11:02
Preparing for cleaning...
Please wait...
Clean washing bath and tray
(5200) 2006-01-09 11:02
Power off for cleaning.
Standby
switch