AX-4030
5-7
AUTION MAX AX-4030 OPERATING MANUAL
5.2.2 Causes and Remedies
Wear protective gloves to prevent exposure to pathogenic microbes before
any operation that may expose you to samples.
Discard used samples, test strips and protective gloves in accordance with
local regulations for biohazardous waste.
E101 Version change
Cause The main ROM was exchanged before the instrument was turned on.
Remedy
Press OK to clear the error.
E102 Power down
Cause The instrument was accidentally turned off during measurement.
Remedy
Press OK to clear the error.
↓
Allow the instrument to warm-up. When the instrument puts itself on standby, retry testing the
remaining samples.
E103 Battery voltage error
Cause The backup battery naturally discharged while the instrument was not in use for several days.
Remedy
Press OK to clear the error.
↓
Keep the instrument powered on for at least 25 hours to charge the battery.
↓
Set the system clock correctly (see page 3-17).
E104 Backup data error
Cause The backup data was corrupted.
Remedy
Press OK to clear the error.
↓
Keep the instrument powered on for at least 25 hours to charge the battery.
↓
Set the system clock correctly (see page 3-17).