2-21
AUTION MAX AX-4030 OPERATING MANUAL
2.3.5 Preparing Samples
Transfer samples into sample tubes, load the tubes into the sample racks, and place the sample
racks in the sampler of the instrument. See “2.2.2 Handling Samples” on page 2-10 to learn
precautions on samples.
Prepare: Sample tubes, item racks, and protective gloves
Wear protective gloves to prevent exposure to pathogenic microbes.
1 Prepare sample tubes.
1 Prepare the sample tubes that comply with the
standards shown at right.
REFERENCE: To have the built-in barcode reader
read barcodes successfully, label the tubes as
shown at right.
2 Pour samples into sample tubes.
1 Fill a sample tube with at least 2 mL of sample.
Ensure the sample level is at least 40 mm above the
bottom of the tube.
IMPORTANT: An insufficient volume of sample
may produce incorrect results.