AX-4030
5-3
AUTION MAX AX-4030 OPERATING MANUAL
5.1.2 Causes and Remedies
Wear protective gloves to prevent exposure to pathogenic microbes before
any operation that may expose you to samples.
Discard used samples, test strips and protective gloves in accordance with
local regulations for biohazardous waste.
NOTE: If a warning occurs during measurement, you will be asked whether to continue the
sample tests or not. Select Continue to measure the remaining samples, and Stop to cancel
the tests.
W001 The feeder has run out of test strips.
Load new strips to continue.
Cause The feeder has run out of test strips.
Remedy Add new test strips to the feeder.
↓
Press
OK to clear the warning.
W002 The STAT port is empty.
Set a sample as explained in the operating manual and try again.
Cause There is no sample in the STAT port, or the sample is incorrectly placed in the STAT port.
Remedy Place the sample in the STAT port correctly (see page 2-27).
↓
Press
OK to clear the warning.
W003 The washing solution has run out.
Add fresh washing solution to continue.
Cause 1. The washing solution bottle is empty.
2. The liquid sensor of the washing solution bottle is not working correctly because the bottle is
uncapped or the liquid sensor cord is disconnected from the instrument.
Remedy 1. Discard the remaining washing solution and add newly prepared solution to the bottle (see
page 4-15).
↓
Press
OK to clear the warning.
2. Connect one end of the liquid sensor cord to the cap of the washing solution bottle, and cap
the bottle. Connect the other end of the cord to the washing solution sensor terminal “S” on
the rear of the instrument (see page 1-27).
↓
Press
OK to clear the warning.