5-20
AX-4030
AUTION MAX AX-4030 OPERATING MANUAL
5.4.3
Turbidity Measurement-Related Errors (Printed Results Report: TURB)
This section lists the error messages that may be added to the “TURB” line on the printed results
reports when incorrect turbidity results are obtained. See the tables below to solve the problem. If
the error persists, contact your distributor.
CAL. ERR
Cause The result obtained from turbidity measurement is outside the specified range.
Remedy Turn off the power, and contact your distributor.
ERROR
Cause Turbidity measurement was not made for some reason.
Remedy Remove the cause.
Retry to measure the sample.
5.4.4 Messages on the Abnormal Results List
If you obtain an abnormal result, see the printed abnormal results list (see pages 2-54 and 3-24) to
determine the cause, and solve the problem by referencing the tables below. The abnormal results
list includes only the errors that occurred after power-on. If an error persists, turn off the power and
contact your distributor.
Wear protective gloves to prevent exposure to pathogenic microbes before
any operation that may expose you to samples.
Discard used samples, test strips and protective gloves in accordance with
local regulations for biohazardous waste.
Barcode misread
Description The barcode could not be read from the sample tube.
Cause 1. The sample tube is not labeled with a barcode, or the barcode is too dirty to read.
2. The barcode is labeled out of position.
3. The built-in barcode reader is out of order.
Remedy 1. Newly label the sample tube, and retry measurement.
2. Reattach the label in the correct position and retry measurement.
3. Turn off the power, and contact your distributor.
No sample
Description No sample was detected.
Cause The sample tube is empty.
Remedy Ensure the sample tube contains at least 2 mL of sample, and the sample level is 40 mm
above the bottom of the tube.
Retry measurement.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Arkray AUTION MAX AX-4030 and is the answer not in the manual?

Arkray AUTION MAX AX-4030 Specifications

General IconGeneral
BrandArkray
ModelAUTION MAX AX-4030
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals