AUTION MAX AX-4030 OPERATING MANUAL
C
Control Measurement2.6
Perform control measurement at regular intervals so as to maintain the status of the instrument and
measurement quality. Up to 3 types or concentrations of controls can be measured by loading them
into the STAT and control rack (control measurement can be performed using at least one control,
and does not always require three different controls). Up to 200 control measurement results can be
stored in memory.
Prepare: Commercially available controls, STAT and control rack, sample tubes,
and protective gloves
Wear protective gloves to prevent exposure to pathogenic microbes.
Discard used controls, test strips, and protective gloves in accordance
with local regulations for biohazardous waste.
REFERENCE: This section describes how to perform control measurement when on standby.
However, controls can also be measured before or in course of normal measurement by
inserting the STAT and control rack at the beginning of or between the racks containing patient
samples.
REFERENCE: For the controls labeled with barcodes, “control IDs” read from the barcodes are
automatically assigned to the corresponding control measurement results.
1 Check the main feeder.
The test strips in the main feeder are always used for control measurement.
1 On the standby screen, check if the main feeder
and the test strips for the feeder are set as you
desire.
The main feeder is indicated by a light blue
background.
2 Load the test strips into the main feeder.
1 Place the test strips in the main feeder.
Any types of test strips can be used as long as they are supported by the instrument.
For detailed instructions, see “2.3.3 Loading Test Strips into the Feeders” on page 2-16. The
main feeder and test strips settings can be changed in step 1 in that section.
Ready for measurement
2006-01-09 13:00
Measurement start No.
0001
Test strips in feeder 1
1: 9EB
Test strips in feeder 2
2: 9EB
Menu Meas No. Strip S.G. Cal
Type of test strips
Main feeder (light blue background)

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Arkray AUTION MAX AX-4030 and is the answer not in the manual?

Arkray AUTION MAX AX-4030 Specifications

General IconGeneral
BrandArkray
ModelAUTION MAX AX-4030
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals