AX-4030
4-37
AUTION MAX AX-4030 OPERATING MANUAL
D
4.5.2 Starting the Instrument Up After Long Periods of Disuse
Use the following procedure to start up the instrument after 1 week or more of disuse.
Prepare: Concentrated washing solution 3 (for AUTION MAX), and purified water
1 Fill the washing solution bottle with the washing solution.
1 Prepare the washing solution and transfer it to the
washing solution bottle.
See “4.3.1 Replacing the Washing Solution” on
page 4-15.
2 Turn on the power.
1 Plug the power cord into an AC receptacle.
2 Press the ON side of the main power switch.
3 Press the standby switch to turn on the power.
During warm-up, the instrument will fill the flow
lines with washing solution.
3 Load the test strips.
1 Make sure the standby screen is displayed.
2 Load the test strips into the feeders.
See “2.3.3 Loading Test Strips into the Feeders” on
page 2-16.
IMPORTANT: The quality of the test strips in
the feeders is ensured for three days. Using
the strips left in the feeders for longer periods
may produce incorrect results.
Concentrated washing
solution 3
Purified water
1800 mL
200 mL
Example: To prepare 2000 mL of ten-fold
diluted washing solution
Standby
switch
ON
OFF
Main power
switch (rear)

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Arkray AUTION MAX AX-4030 and is the answer not in the manual?

Arkray AUTION MAX AX-4030 Specifications

General IconGeneral
BrandArkray
ModelAUTION MAX AX-4030
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals