AUTION MAX AX-4030 OPERATING MANUAL
1.4.5 Installing the Drain Bottle
Set the drain bottle at the same or a lower height than the surface that the
instrument sits on. If higher than the instrument base, liquid waste may
not drain properly and consequently leak. Contacting liquid waste with
unprotected hands may expose you to pathogenic microbes.
Urine samples and solution are discharged into the drain bottle. Place the bottle on a flat, level surface.
Prepare: Drain bottle and silicon tube [3 mm (i.d.) × 6 mm (o.d.), 4 m]
1 Connect the tube to the drain bottle.
1 Cut the silicon tube to an appropriate length.
2 Remove the joint from the drain bottle. Insert one
end of the silicon tube into the joint until the tube
extends about 5 cm from the joint.
3 Reattach the joint to the bottle cap, and tighten it
securely.
2 Connect the tube to the instrument.
Do not bend or pinch the silicon tube of the drain bottle. Also, keep
objects off of the tube. If liquid waste clogs inside the tube, the tube may
disconnect from the instrument and spill contained liquid. Contacting liquid
waste with unprotected hands may expose you to pathogenic microbes.
1 Fit the other end of the tube over the drain joint “D” on the rear panel.