4-26
AX-4030
AUTION MAX AX-4030 OPERATING MANUAL
1 Press the “PUSH” mark on the STAT port
backward to unlock the stopper, and then slide the
port toward you.
2 Take the sample tube out of the STAT port.
4.4.4 Cleaning the Washing Bath
The washing bath becomes contaminated as more urine tests are run. Clean the washing bath at
least once a month or when the washing bath looks seriously contaminated.
Prepare: Purified water, cotton swabs, and protective gloves
Wear protective gloves to prevent exposure to pathogenic microbes.
Discard used cleaning tools and protective gloves in accordance with local
regulations for biohazardous waste.
1 Move the nozzle.
Move the nozzle to make enough room for you to access the washing bath.
1 On the standby screen, press Menu to go to the
[Main menu] screen.
2 Press 5 to go to the [Maintenance] screen.
PUSH
Menu Meas No. Strip S.G. Cal
Main menu
(0000) 2006-01-09 11:00
1 Parameter settings
2 Results
3 Initial settings
4 Print
5 Maintenance
Go back

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Arkray AUTION MAX AX-4030 and is the answer not in the manual?

Arkray AUTION MAX AX-4030 Specifications

General IconGeneral
BrandArkray
ModelAUTION MAX AX-4030
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals