EasyManuals Logo

Arkray AUTION MAX AX-4030 User Manual

Arkray AUTION MAX AX-4030
206 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #177 background imageLoading...
Page #177 background image
4-34
AX-4030
AUTION MAX AX-4030 OPERATING MANUAL
D
4.5.1 Preparing the Instrument Before Long Periods of Disuse
When you intend not to use the instrument for a week or more, use the procedure below to clean
each section. Failure to do that can cause fluid remaining in the flow lines to crystallize and
consequently damage the instrument.
Prepare: 0.5% sodium hypochlorite solution (or commercially available detergent diluted
appropriately), sample tube, cotton swabs, purified water, alcohol, blower brush, tissue
paper, and protective gloves
Wear protective gloves to prevent exposure to pathogenic microbes.
Discard used cleaning tools and protective gloves in accordance with local
regulations for biohazardous waste.
1 Wash the S.G. cell.
1 Wash the S.G. cell.
See “4.4.3 Cleaning the S.G. Cell” on page 4-24.
Keep the power on after the completion of the washout.
2 Clean the washing bath.
1 Clean the washing bath.
See “4.4.4 Cleaning the Washing Bath” on page 4-26.
3 Flush the flow lines.
1 Make sure the standby screen is displayed, then
press the standby switch to turn off the power.
Standby
switch
Before/After Long Periods of Disuse4.5

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Arkray AUTION MAX AX-4030 and is the answer not in the manual?

Arkray AUTION MAX AX-4030 Specifications

General IconGeneral
BrandArkray
ModelAUTION MAX AX-4030
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals