1.6.1 Precautions in Instrument Relocation
Before relocating the instrument, read the following instructions and always take proper safety
precautions.
Remove the drain bottle, washing solution bottle and sampler from the
instrument.
Drain washing solution from the flow line. Moving the instrument with washing
solution in the flow line may damage the instrument.
Turn off the power by pressing the standby switch followed by the OFF side of the
main power switch. Then, disconnect the power cord.
Disconnect the optional hand-held barcode reader and the communication cables
for external devices, if any, from the rear panel of the instrument.
Make sure the front, maintenance and side covers are closed before relocating
the instrument. Moving the instrument with any of the covers open may expose
you to pathogenic microbes or damage the instrument.
Make sure there are no test strips left inside the instrument before relocating the
instrument. Moving the instrument with used test strips inside may expose you to
pathogenic microbes or damage the instrument.
For safety reasons, always move the instrument with the help of at least one other
person. Hold the bottom of the instrument with both hands and be careful not to
impact or shake the instrument. Rough handling may damage the instrument.
Read
“1.4.1 Precautions in Instrument Installation” on page 1-18 before relocating
the instrument.
1.6.2 Checking for Leftover Test Strips
Prepare: Protective gloves
Wear protective gloves to prevent exposure to pathogenic microbes.
Discard used protective gloves, test strips and liquid waste in accordance
with local regulations for biohazardous waste.
1 Check for test strips in the feeders.
1 Make sure the standby screen is displayed.
2 Open the 2 feeder covers.
Relocation1.6

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Arkray AUTION MAX AX-4030 and is the answer not in the manual?

Arkray AUTION MAX AX-4030 Specifications

General IconGeneral
BrandArkray
ModelAUTION MAX AX-4030
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals