AX-4030
4-25
AUTION MAX AX-4030 OPERATING MANUAL
3 Press 1 to go to the [S.G. cell washout] screen.
4 Press Start to start washing.
During S.G. cell washout, the remaining time before
completion is displayed.
REFERENCE: If warning “W002” occurs
The sample tube containing the detergent may be
incorrectly placed in the STAT port. Load the tube
correctly, and press OK to clear the warning.
REFERENCE: To stop washout
Press Stop . The instrument will flush the flow
lines, and displays the [Maintenance] screen
again.
3 When washout is completed…
The end process will automatically start, and then
the [Maintenance] screen will appear again.
Press Go back twice on the [Maintenance] screen to
return to the standby screen.
Maintenance
(5000) 2006-01-09 11:00
1 S.G. cell washout
2 Clean washing bath and tray
3 Check measurement
4 Maintenance information
S.G. cell washout
(5100) 2006-01-09 11:02
Set washing solution in STAT port.
Start Go back
S.G. cell washout
(5100) 2006-01-09 11:03
Washing...
Please wait...
5'00
S.G. cell washout
(5100) 2006-01-09 11:08
S.G. cell washout ending.
Please wait...

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Arkray AUTION MAX AX-4030 and is the answer not in the manual?

Arkray AUTION MAX AX-4030 Specifications

General IconGeneral
BrandArkray
ModelAUTION MAX AX-4030
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals