AX-4030
2-37
AUTION MAX AX-4030 OPERATING MANUAL
3 Start the S.G. calibration.
1 Make sure the sample rack with the S.G. standard
solutions are correctly loaded into the sampler.
2 Press Start to start calibration.
• Measurement of the standard solutions will take
place when the sample rack reaches the aspirating
position.
REFERENCE: To stop calibration in progress,
press Stop . The instrument will wait until the
solution measurement in progress completes and
then proceeds to the breaking process before
putting itself back on standby. Note that breaking
calibration aborts all of the calibration results.
4 When S.G. calibration is completed…
• The message “Measurement completed” will
appear.
1 Press Go back to return to the standby screen.
2 Remove the sample rack of the standard solutions
from the sampler.
REFERENCE: If the error “E120” occurs
See “E120” in “5.2.2 Causes and Remedies” on page 5-8 to determine the cause of the error,
and retry from step 1 . If the error persists, there is something wrong with the instrument.
Contact your distributor.