4-4
AX-4030
AUTION MAX AX-4030 OPERATING MANUAL
2 Sterilize the waste box.
1 Sterilize the waste box with alcohol.
2 Rinse the box with water to remove dirt.
3 Wipe water off the surface.
3 Install the waste box.
1 Hold the waste box tab facing forward, and place
the box on the waste box tray.
Ensure the waste box guides fit the concaves inside
the tray
2 Store the waste box tray in the instrument.
If the waste box is correctly set in the tray, it is held
steadily by the magnets.
If the waste box gets caught halfway, it is set
incorrectly.
NOTE: Ensure the waste box correctly fits into the
tray before pushing it in. Incorrect installation may
cause used test strips to be scattered around
inside the instrument or to clog the test strip path.
One guide
Tab
Two guides