T210 Test strip introduction section drive trouble
Cause 1. The transport tray or introduction tray is not installed correctly.
2. The test strips cannot be transported because the transport section is dirty or obstructed.
Remedy
1. Press OK to clear the trouble.
↓
Install the tray correctly (see pages 4-12 or 4-19).
2. Press OK to clear the trouble.
↓
Clean the dirty transport tray, or remove the obstacles, if there are any.
T211 Test strip transport section drive trouble
Cause 1. The waste box is overflowing with used test strips.
2. The transport section did not work correctly.
Remedy
1. Press OK to clear the trouble.
↓
Pull out the waste box and discard the used test strips (see page 4-3).
↓
Remove the test strips scattered on the analysis section.
2. Press OK to clear the trouble.
↓
Check the test strip transport section. If the section is damaged, turn off the power and
contact your distributor.
T212 Test strip identification section drive trouble
Cause The test strip flip section of the feeder is clogged with test strips.
Remedy
Press OK to clear the trouble.
↓
Remove the clogged test strips from the feeder.
T220 Optical unit drive trouble
Cause 1. The transport tray is not installed correctly.
2. The waste box is overflowing with used test strips that are blocking the measurement light
and the optical unit movement.
3. Trouble occurred in the optical unit.
Remedy
1. Press OK to clear the trouble.
↓
Install the transport tray correctly (see page 4-19).
2. Press OK to clear the trouble.
↓
Pull out the waste box and discard the used test strips (see page 4-3).
3. Press OK to clear the trouble.
↓
Check the optical block. Remove the test strips that are blocking movement. If the optical
block is damaged, turn off the power and contact your distributor.
↓
Remove the test strips scattered on the analysis section.