AVEA ventilator systems Chapter 6: Infant Non-invasive Ventilation 161
L2786 Rev. M
4. Touch Continue; the following message appears.
If the calibration is successful, the following message appears.
Note:
If the calibration test fails, check the following:
• Ensure the patient was disconnected during the calibration.
1. Ensure the circuit connections are secure.
2. Ensure there was no movement of the circuit during the calibration.
3. Ensure the prongs are open during the test.
4. Ensure the expiratory limb of the circuit was disconnected before starting the calibration.
If failure of the calibration persists after checking all of the above, remove the ventilator from service and have it
checked by a qualified technician.
5. Leave the patient disconnected and reconnect the expiratory limb of the circuit at the patient wye.