F15 Series Fetal & Maternal Monitor User Manual After Monitoring
- 110 -
NOTE:
The monitor checks uterine contract (UC) pressure when the NIBP key is pressed. If the UC is
over 50, a prompt “Intense UC, can't measure NIBP now.” is issued, and the monitor will check
the UC every 30 seconds. The monitor will measure NIBP only when the UC is lower than 50,
and it will then start timing for the Auto measurement.
To stop the current measurement,
Press the NIBP key anytime during the current measurement to stop it. Another measurement will
start after the time interval.
WARNING
Prolonged NIBP measurements in automatic mode may be associated with purplish patches,
ischemia and neurologic damage in the limb wearing the cuff. When monitoring a patient,
examine the extremities of the limb frequently for normal color, warmth and sensitivity. If any
abnormality is observed, stop the NIBP measurement.
8.4.5 *Manual Measurement
To perform a manual measurement,
1 Select the main menu key on the main interface.
2 Select Mother > Cycle.
3 Select Manual.
4 Select OK.
5 Press NIBP key on the main interface to start a manual measurement.
To stop the manual measurement,
Press the NIBP key anytime during the measurement to stop it.
To perform a manual measurement during an auto measurement interval,
1 Press the NIBP key to start the manual measurement.
2 Press the NIBP key again anytime to stop it.
The monitor will restart timing for the Auto measurement and resume measuring after the time interval.
NOTE:
1 If you are in doubt about the accuracy of any reading(s), check the patient's vital signs by
an alternative method before checking the functioning of the monitor.
2 The monitor checks uterine contract (UC) pressure when the NIBP key is pressed. If the UC
is over 50, a prompt “Intense UC, can't measure NIBP now.” is issued. Please wait and do
not attempt to measure NIBP until the UC is lower than 50.
CAUTION
1 Do not squeeze the rubber tube on the cuff.
2 If liquid is inadvertently splashed on the equipment or its accessories, or may enter the
conduit or inside the monitor, contact local service center.
8.4.6 Changing NIBP Unit
You can change the NIBP unit.
1 Select the main menu key on the main interface.
2 Select Mother > Unit (NIBP Setup).
3 Select mmHg (default) or kPa.
Traditional 510(k) of Fetal & Maternal Monitor