F15 Series Fetal & Maternal Monitor User Manual Alarms
- 62 -
Probe MALF (FHR3)
US transducer 3 is
disconnected with the monitor
Connect US transducer 3 with the
monitor
Probe MALF (DECG)
DECG&IUP transducer is
disconnected with the monitor
Connect DECG&IUP transducer with the
monitor
Probe MALF (TOCO)
TOCO transducer is
disconnected with the monitor
Connect TOCO transducer with the
monitor
Probe MALF (FECG)
FECG Fetal&Maternal Module
is disconnected with the
monitor
Connect FECG Fetal&Maternal Module
with the monitor
Current network
environment poor
(US1)
The current network
environment is poor and US
transducer 1 has problem to
connect with the monitor
Check whether the ambulating mother
has reached the limit of the active signal
area-of-reach and whether the network
communication is fine.
Current network
environment poor
(US2)
The current network
environment is poor and US
transducer 2 has problem to
connect with the monitor
Check whether the ambulating mother
has reached the limit of the active signal
area-of-reach and whether the network
communication is fine.
Current network
environment poor
(US3)
The current network
environment is poor and US
transducer 3 has problem to
connect with the monitor
Check whether the ambulating mother
has reached the limit of the active signal
area-of-reach and whether the network
communication is fine.
Current network
environment poor
(TOCO)
The current network
environment is poor and TOCO
transducer has problem to
connect with the monitor
Check whether the ambulating mother
has reached the limit of the active signal
area-of-reach and whether the network
communication is fine.
Current network
environment poor
(DECG)
The current network
environment is poor and
DECG&IUP transducer has
problem to connect with the
monitor
Check whether the ambulating mother
has reached the limit of the active signal
area-of-reach and whether the network
communication is fine.
Current network
environment poor
(FECG)
The current network
environment is poor and FECG
Fetal&Maternal Module has
problem to connect with the
monitor
Check whether the ambulating mother
has reached the limit of the active signal
area-of-reach and whether the network
communication is fine.
Network is offline
The central station disconnects
with the fetal monitor
(ETHERNET network)
Examine the network connection
between the central station and the fetal
monitor
4.18.2 Maternal Monitoring Alarm Messages
Besides the fetal monitoring alarms, F15 also gives alarms for the situations that occur during
maternal monitoring. The alarm messages are listed below.
Traditional 510(k) of Fetal & Maternal Monitor
014_14.1_F15_Series_User_Manual
78202

Table of Contents

Other manuals for EDAN F15 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the EDAN F15 Series and is the answer not in the manual?

EDAN F15 Series Specifications

General IconGeneral
BrandEDAN
ModelF15 Series
CategoryMonitor
LanguageEnglish

Related product manuals