F15 Series Fetal & Maternal Monitor User Manual Alarms
- 63 -
Patient Alarm Messages
Alarm Message Cause Countermeasure
High Priority
***ASYSTOLE
No QRS wave is detected by
TOCO&MECG transducer in 4
seconds or no QRS wave is
detected by FECG
Fetal&
Maternal Module in 6
seconds
Check the patient’s condition and take
necessary measures.
Medium Priority
**HR HIGH or
**HR xxx > yyy
Maternal HR result (xxx) is higher
than the upper limit (yyy).
Check if the alarm limits are suitable;
check the patient’s condition.
**HR LOW or
**HR xxx < yyy
Maternal HR result (xxx) is lower
than the lower limit (yyy).
** SpO
2
HIGH or
** SpO
2
xxx > yyy
SpO
2
result (xxx) is higher than
the upper limit (yyy).
2
** SpO
2
xxx < yyy
2
result (xxx) is lower than the
lower limit (yyy).
**SYS HIGH or
**SYS xxx > yyy
SYS result (xxx) is higher than the
upper limit (yyy).
**SYS LOW or
**SYS xxx < yyy
SYS result (xxx) is lower than the
lower limit (yyy).
**DIA HIGH or
**DIA xxx > yyy
DIA result (xxx) is higher than the
upper limit (yyy).
**DIA LOW or
**DIA xxx < yyy
DIA result (xxx) is lower than the
lower limit (yyy).
**MAP HIGH or
**MAP xxx > yyy
MAP result (xxx) is higher than the
upper limit (yyy).
**MAPLOW or
**MAP xxx < yyy
MAP result (xxx) is lower than the
lower limit (yyy).
**TEMP xxx > yyy
TEMP result (xxx) is higher than
the upper limit (yyy).
**TEMP LOW or
**TEMP xxx < yyy
TEMP result (xxx) is lower than
the lower limit (yyy).
Technical Alarm Messages
Alarm Message Cause Countermeasure
High Priority
***
high
CPU temperature exceeds the
safe limits
Check whether the fan of the monitor
works normally, if no, please contact the
manufacturer for service