F15 Series Fetal & Maternal Monitor User Manual After Monitoring
- 111 -
4 Select OK.
Note:
When the network version is ETHERNET 1.4 and above,if the fetal monitor gets on line, the
central station will control to make the NIBP unit of the fetal monitor in concert with that of the
central station.
8.4.7 *Choosing NIBP Printing Mode
When the recorder is printing real-time fetal traces, the NIBP result is also recorded on the paper
whenever NIBP measurement is performed. After the paper stops advancing, you can choose to keep
recording NIBP results on the paper.
To enable or disable NIBP printing after paper advancing stops,
1 Select the main menu key on the main interface.
2 Select Recorder > NIBP.
3 Select ON or OFF (default).
4 Select OK.
NOTE:
The NIBP measurement results during the period of paper lacking and fast printing after new
paper is loaded will not be printed. Therefore, do not perform NIBP measurements during this
period.
8.4.8 *Calibrating NIBP
NIBP is not user-calibrated. Cuff-pressure transducers must be verified and calibrated, if
necessary, at least once every two years by a qualified service professional.
8.5 Maternal TEMP Monitoring
8.5.1 Monitoring Procedure
1) Insert the TEMP plug into the TEMP socket on the monitor.
2) Apply the sensor firmly underneath the patient’s axilla. It takes 5 minutes for the temperature
measurement to stabilize.
WARNING
1 Check if the TEMP sensor functions properly prior to use.
2 Do not apply the TEMP sensor to the mouth or the rectum.