F15 Series Fetal & Maternal Monitor User Manual Maintenance and Cleaning
- 128 -
7. Inspect the transducer to ensure that there is no damage.
11.3.2.2 Disinfecting the NIBP Cuff
Disinfecting the Cuff:
1. Take out the air bladder before disinfection.
2. Wipe the cuff and the air bladder with a soft cloth dampened with the disinfectant solution.
3. Leave the cuff and air bladder to air dry for at least 30 minutes.
Replacing the Air Bladder:
After disinfection, replace the air bladder into the cuff. Refer to Section 11.2.2.2 for more information.
NOTE:
Prolonged use of disinfectant may cause discoloration of the cuff.
11.3.2.3 Disinfecting the ECG Cable Assembly
1. Wipe the cable assembly with a soft cloth dampened with the disinfectant solution.
Wipe off the disinfectant solution with a dry cloth after disinfection.
2. Leave the cable assembly to air dry for at least 30 minutes.
11.3.2.4 Disinfecting the TOCO Transducer and SpO2 Sensor
1. Wipe the surfaces of the sensor and cable using a soft cloth dampened with the disinfection
solution.
2. Wipe the patient contact area of the sensor with the cotton swab dampened with the disinfection
solution.
3. Wipe off the disinfection solution with a dry cloth after disinfection.
4. Leave the sensor to air dry for at least 30 minutes.
11.3.2.5 Disinfecting the TEMP Sensor
1. Wipe the patient contact area with a soft cloth dampened with the disinfectant solution (ethanol or
isopropanol).
2. Wipe off the disinfectant solution with a dry cloth after disinfection.
3. Leave the sensor to air dry.
11.3.3 Cleaning and Disinfecting Other Accessories
For cleaning and disinfecting other accessories, refer to the instructions delivered with the
accessories. If the accessories are not accompanied by instructions, refer to this manual for the
methods of cleaning and disinfecting the monitor.
11.4 Storage
1. Ensure the monitor and reusable accessories are cleaned, disinfected and completely dried
before storage.
2. Store the monitor and reusable accessories in a dry environment free of dust.
3. Store the monitor and reusable accessories under the following conditions: