F15 Series Fetal & Maternal Monitor User Manual Alarms
- 53 -
4 Select Alarm Pause (default) or Alarm Reset.
If Alarm Pause is selected: When the monitor gives out alarm sound and you press the
SILENCE key, the alarm indicator displays , and the alarm sound is muted.
If Alarm Reset is selected: When the monitor gives out alarm sound and you press the
SILENCE key, the alarm indicator displays , and the alarm sound is muted. When
other alarms present, the monitor will enable the audible alarm again automatically.
5 Select OK.
4.10 To Switch Individual Measurement Alarms On or Off
1 Select the main menu key on the main interface.
2 Select Alarm. On the displayed Password box, enter user password, and then select
Enter.
3 Select a measurement parameter and the corresponding alarm setup menu pops up.
4 Select ON (default) or OFF.
5 Select OK.
6 If the alarm is switched off, an alarm switched-off symbol will appear in the numeric
window; if the alarm is switched on, the symbol will disappear.
WARNING
Do not switch the alarm off for the condition where the patient’s safety maybe endangered.
4.11 Changing the Alarm Limits
The alarm limits you set determine the conditions that trigger the alarm.
1 Select the main menu key on the main interface.
2 Select Alarm. On the displayed Password box, enter user password, then select Enter.
3 Select a measurement parameter and the corresponding alarm setup menu pops up.
4 Select a value from High Alarm Limit.
5 Select a value from Low Alarm Limit.
6 Select OK.
4.12 Switching Transducer Unplugged On or Off
You can switch “Transducer Unplugged” on or off:
1 Select the main menu key on the main interface.
2 Select Alarm. On the displayed Password box, enter user password, and then select Enter.
3 Select Transducer unplugged on the alarm settings interface.
4 Select ON or OFF (default).
5 Select OK.
Traditional 510(k) of Fetal & Maternal Monitor
014_14.1_F15_Series_User_Manual