F15 Series Fetal & Maternal Monitor User Manual Basic Operation
- 48 -
2) CTG Analysis Results:
SIGNAL LOSS:
The proportion of the signal loss. If it is larger than 10%, analysis
results cannot be acquired.
CONTRACTIONS:
The contraction time during analysis.
BASAL HEART
RATE:
The average FHR in 10 minutes when it is not influenced by fetal
movement or contractions.
ACCELERATIONS:
The acceleration time, including the acceleration with amplitude larger
than 10bpm and lasts more than 10 seconds, and the acceleration with
amplitude larger than 15bpm and lasts more than 15 seconds.
DECELERATIONS:
The deceleration time.
SHORT TERM
VARIATION:
The short-term variation analysis result.
LONG TERM
VARIATION:
The long-term variation analysis result.
ANALYSIS START:
The start time of the analysis.
ANALYSIS END:
The finishing time of the analysis.
During 10 to 60-minute of the timer, the monitor gives CTG analysis results every minute.
At the end of the printing, the recorder prints the CTG analysis results of this moment on the
recorder paper.
Be aware that CTG analysis result is a calculation output. It can be used as a reference to assist
medical personnel in making correct diagnosis, instead of replacing it.
NOTE:
Do not disconnect the ultrasound transducer(s) before the printing stops, otherwise the
analysis results will not be printed.
3.6.3.6 *Marking a Note
When there is a significant event, you can press the MARK key on the main interface to add a note.
An event mark will appear on both the main interface and the recorder paper.
However, an event mark cannot clearly indicate an event. Smart Notes provides a list of annotation
for the events, including events that relate to drugs, positions, membranes, procedures, antenatal,
reasons and others.
To enable or disable Smart Notes,
1 Select the main menu key on the main interface.
2 Select General > Smart Notes.
3 Select ON or OFF (default).
Once Smart Notes is enabled, press the MARK key on the main interface to open the smart note list,
choose an event catalog and then choose an annotation from the list.
The annotation of this event will be printed in the top area of recorder paper during real-time printing.
You can change the smart note content in the smart note list by performing the following steps:
1 Select the main menu key on the main interface.
2 Select General.
3 Select the smart note editing key .
4 Select a catalog.
5 Select a note.
Traditional 510(k) of Fetal & Maternal Monitor
014_14.1_F15_Series_User_Manual
64202

Table of Contents

Other manuals for EDAN F15 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the EDAN F15 Series and is the answer not in the manual?

EDAN F15 Series Specifications

General IconGeneral
BrandEDAN
ModelF15 Series
CategoryMonitor
LanguageEnglish

Related product manuals