the high-frequency signals, so fetal heart sound is heard. On the other hand, the autocorrelation
algorithm computes the stronger signal source and thereby the maternal heart/pulse rate is displayed.
As a result, the audio output differs from the screen reading.
If this situation occurs, it can be dismissed by repositioning the transducer.
In a word, the abnormalities listed above (artifacts, sound and reading differences) are caused by the
limitations of ultrasonic monitoring technique. Fortunately they rarely occur. But a good understanding
of how to detect them and what countermeasures should be taken will help obtain better fetal
monitoring effect.