F15 Series Fetal & Maternal Monitor User Manual WI-FI Networking and Instructions
- 174 -
laptops with WirelessMon or Android phones with inSSIDer software.
After arriving at the construction site, the sales or after-sales engineer immediately contact the
relevant person in charge of the hospital to communicate and confirm the actual signal coverage, the
placement of the monitor, the information on the number of co-frequency signal devices in the vicinity,
and finally summarize the information to customer service engineer of domestic/international
marketing for discharging. .
Survey the following items:
1) The overall building (room layout).
-Building floor plan/size, room size
- Take a picture of the fire evacuation plan of each floor of the hospital
- The location and size of the ward’s toilet should be indicated on the floor plan
- load-bearing wall/beam, wall thickness, walkway corner
- Decoration materials (metal, glass, brick, wood, etc.)
- ceiling structure
- room use and functional identification of each area
The picture of above information is best taken on-site to the product engineer for reference.
- Finally output the floor plan and wall thickness of the entire ward.
2) Signal coverage area
3) Placement of the central monitoring system, each F-series fetal monitor fixed location and its
moving area.
4) The number of wireless networks in the coverage area.
5) Power supply conditions.
6) Number of devices and installation location.
7) Contact information of users and agents.
Before the start of construction, negotiate the construction time limit and the issue of punching through
the wall with the relevant person in charge of the hospital. Based on the above survey results, submit
the network plan and network requirements table.
Traditional 510(k) of Fetal & Maternal Monitor
014_14.1_F15_Series_User_Manual