F15 Series Fetal & Maternal Monitor User Manual Alarms
- 56 -
FHR1/DFHR/FHR2/FHR3 Low
Alarm Limit
60 bpm ~ 205 bpm, in increments of 5
110 bpm
FHR1/DFHR/FHR2/FHR3 High
Alarm Limit
65 bpm ~ 210 bpm, in increments of 5
160 bpm
FHR1/DFHR/FHR2/FHR3 High
Alarm Delay
0 ~ 300 second(s), in increments of 5
10 seconds
FHR1/DFHR/FHR2/FHR3 Low
Alarm Delay
0 ~ 300 second(s), in increments of 5
10 seconds
FHR1/DFHR/FHR2/FHR3
Alarm Priority
Medium, not adjustable
Medium
HR Alarm
On, Off
On
HR Low Alarm Limit
28 bpm ~ 242 bpm, in increments of 1
50 bpm
HR High Alarm Limit
29 bpm ~ 243 bpm, in increments of 1
120 bpm
HR Alarm Delay
0 second, not adjustable
0 second
HR Alarm Priority
Medium, not adjustable
Medium
SpO
2
Alarm
On, Off
On
SpO
2
Low Alarm Limit
50% ~ 99%, in increments of 1
90%
SpO
2
High Alarm Limit
51% ~ 100%, in increments of 1
100%
SpO
2
Alarm Delay
0 second, not adjustable
0 second
SpO
2
Alarm Priority
Medium, not adjustable
Medium
SYS Alarm
On, Off
On
SYS Low Alarm Limit
40 mmHg ~ 269 mmHg, in increments of 1
(5.3 kPa ~ 35.9 kPa, in increments of 0.1)
90 mmHg
(12.0 kPa)
SYS High Alarm Limit
41 mmHg ~ 270 mmHg, in increments of 1
(5.4 kPa ~ 36.0 kPa, in increments of 0.1)
160 mmHg
(21.3 kPa)
SYS Alarm Delay
0 second, not adjustable
0 second
SYS Alarm Priority
Medium, not adjustable
Medium
DIA Alarm
On, Off
On
DIA Low Alarm Limit
10 mmHg ~ 214 mmHg, in increments of 1
(1.3 kPa ~ 28.6 kPa, in increments of 0.1)
50 mmHg
(6.8 kPa)
DIA High Alarm Limit
11 mmHg ~ 215 mmHg, in increments of 1
(1.4 kPa ~28.7 kPa, in increments of 0.1)
90 mmHg
(12.0 kPa)
DIA Alarm Delay
0 second, not adjustable
0 second
DIA Alarm Priority
Medium, not adjustable
Medium
MAP Alarm
On, Off
On
MAP Low Alarm Limit
20 mmHg ~ 234 mmHg, in increments of 1
(2.7 kPa ~31.2 kPa, in increments of 0.1)
60 mmHg
(8.0 kPa)
MAP High Alarm Limit
21 mmHg ~ 235 mmHg, in increments of 1
(2.8 kPa ~31.3 kPa, in increments of 0.1)
110 mmHg
(14.8 kPa)
Traditional 510(k) of Fetal & Maternal Monitor
014_14.1_F15_Series_User_Manual
72202

Table of Contents

Other manuals for EDAN F15 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the EDAN F15 Series and is the answer not in the manual?

EDAN F15 Series Specifications

General IconGeneral
BrandEDAN
ModelF15 Series
CategoryMonitor
LanguageEnglish

Related product manuals