ECI Multi Functional Weighing Controller Operating Manual
Page 141
8.1 Load Calibration
It is the operation of identification of a mass with a known weight to the device. It must be made by using real
weights.
1) It is entered to menu 2.5 for the load calibration.
2) Load cell type is selected. The options are;
1) Esit (2 mV/V)
2) 2.5mV/V
3) 5.0mV/V
4) 10.0mV/V
5) 20.0mV/V
6) 160.0mV/V
mV/V selection is made depending on the load cell model.
3) For entering to Calibration Type menu is pressed.