ECI Multi Functional Weighing Controller Operating Manual
Page 295
Type
No
Short Name
Explanation
Error 1 Maximum Load!
This message is displayed when the measured value is more than
"Capacity" + 9e.
Error 3 Zero Limit!
The load with this value cannot be zerod. It can be zeroed with zero
calibration.
Error
5
LoadCellError
Load cell may be defected or there is a problem with the connection.
Error
6
ADC Data Error!
ADC Module is defective. Consult ESIT.
Error 10 Coeff.OutofLimit
It shows that the coefficient calculated as the result of the calibration is out of
limit. The load used must be increased or the value to be calibrated must be
decreased.
Error 11 Insufficient Load!
It shows that the load used for the calibration is not sufficient. The load used
must be increased.
Error
12
2mV/V Only
It can be used only for 2mV/V load cell.
Error
21
No record!
No back-up has been made for the load cell. Thus it cannot be recovered.
Error
22
Memory 1 Error
The memory unit could not read/write. Consult ESIT.
Error 23 Wrong Value
The parameter attempted to be recorded is out of the allowed limits. Check
the values
Error
24
Memory 2 Error
The memory unit could not read/write. Consult ESIT.
17 Error, Warning and Information Messages