Error 1 Maximum Load!
This message is displayed when the measured value is more than
"Capacity" + 9e.
Error 3 Zero Limit!
The load with this value cannot be zerod. It can be zeroed with zero
calibration.
Load cell may be defected or there is a problem with the connection.
ADC Module is defective. Consult ESIT.
Error 10 Coeff.OutofLimit
It shows that the coefficient calculated as the result of the calibration is out of
limit. The load used must be increased or the value to be calibrated must be
Error 11 Insufficient Load!
It shows that the load used for the calibration is not sufficient. The load used
must be increased.
It can be used only for 2mV/V load cell.
No back-up has been made for the load cell. Thus it cannot be recovered.
The memory unit could not read/write. Consult ESIT.
Error 23 Wrong Value
The parameter attempted to be recorded is out of the allowed limits. Check
the values
The memory unit could not read/write. Consult ESIT.