ECI Multi Functional Weighing Controller Operating Manual
Page 296
Error
31
PrinterTime.Out
Printer Connection is wrong or there is no printer connection.
Error
42
Battery Error
Date/Time info is unable to be read. Change the battery.
Error
50
No USB
USB is not inserted or detected.
Error
51
No File
No File in USB
Error
52
Corrupted File
Reload the file
Error
53
Restart!
Shut-down and turn on the device.
Error 60
SensitiveQuant.Erro
r
Sensitive Quantity cannot be more than the Target Value. (Shot Value +
Sensitive Quantity value) also cannot be more than the Target Value.
Error 61 ShotValueError
Shot Value cannot be more than the Target Value. (Shot Value + Sensitive
Quantity value) also cannot be more than the Target Value.
Error 62 LowTareValue
The operation cannot be started as the Tare value is lower than the value
entered to the Minimum Tare menu.
Error 63 HighTareValue
The operation cannot be started as the Tare value is higher than the value
entered to the Minimum Tare menu.
Error
64
TargetValue.Excess
Target Value cannot be more than the Capacity Value.
Error 66 Scale is full
The filling operation could not be started as the target weight is less than the
the weight on the weighing platform. (When Auto Tare is Off)
Error
70
InsufficientDigit
The digit is not sufficient for showing the value measured.
Error
90
ADC ModulError
ADC could not be read. Consult ESIT.
Error
91
Defective Sensor
The temperature sensor could not be read. Consult ESIT.
Warning
101
Mobility
Zeroing or taring could not be performed due to mobility.