ECI Multi Functional Weighing Controller Operating Manual
Page 296
Printer Connection is wrong or there is no printer connection.
Date/Time info is unable to be read. Change the battery.
USB is not inserted or detected.
Shut-down and turn on the device.
Error 60
Sensitive Quantity cannot be more than the Target Value. (Shot Value +
Sensitive Quantity value) also cannot be more than the Target Value.
Error 61 ShotValueError
Shot Value cannot be more than the Target Value. (Shot Value + Sensitive
Quantity value) also cannot be more than the Target Value.
Error 62 LowTareValue
The operation cannot be started as the Tare value is lower than the value
entered to the Minimum Tare menu.
Error 63 HighTareValue
The operation cannot be started as the Tare value is higher than the value
entered to the Minimum Tare menu.
Target Value cannot be more than the Capacity Value.
Error 66 Scale is full
The filling operation could not be started as the target weight is less than the
the weight on the weighing platform. (When Auto Tare is Off)
The digit is not sufficient for showing the value measured.
ADC could not be read. Consult ESIT.
The temperature sensor could not be read. Consult ESIT.
Zeroing or taring could not be performed due to mobility.