Test Results
Power Blazer 625
Summary - Time Error / Wander
Time Error / Time Interval Error Metrics
MTIE label and the pass/fail verdict icon are displayed when enabled
and the test is running.
TDEV label and the pass/fail verdict icon are displayed when enabled
and the test is running.
For Time Error measurement with either G.8275.1 profile or 1PPS interface:
Note: The following metrics are collected using 0.1 Hz Low-Pass filtering.
TE (ns) indicates the last time error value collected in the last second.
Max TE (ns) indicates the maximum time error value collected since
the beginning of the test.
Min TE (ns) indicates the minimum time error value collected since
the beginning of the test.
Max Absolute TE (ns) indicates the maximum absolute time error
value collected since the beginning of the test. The pass/fail verdict is
also displayed when enabled.
dTE
H
pk-pk (ns) indicates the dynamic time error peak-to-peak over a
defined observation period (refer to dTE
H
Observation Period on
page 448). The first value is only available after the initial obsersation
period has elapsed, the value is then updated every second.
Max dTE
H
pk-pk (ns) indicates maximum dynamic time error
peak-to-peak since the beginning of the test. The pass/fail verdict is
also displayed when enabled.