5-7
Chapter 5 Initial Settings Alarm Related Setup
set to OFF.
3 Select the parameters to be monitored during Oxygenator Mode.
Unselected parameters will be displayed with decreased brightness on the home display.
If [Sel. Parameter] is set for "Alarm Setting", the alarm for unselected parameter will not be generated.
4 Set the display mode (Sub Mode/Extended Display 1/Extended Display 2) during Oxygenator Mode.
If [OFF] is selected, the display mode will not switch during the oxygenator mode.
When returning to the standard monitoring mode, it will return to the display mode used before switching
to the Oxygenator Mode.
5 Select whether to enable or disable the <ECG Check Electrodes> alarm and <SpO
2
Check Sensor Attach.>
alarm during oxygenator mode.
If [Disable] is selected, alarms will not generate during the oxygenator mode.
6Set the "HR/PR Lower Limit during Alarm Auto Setting".
None: No limit will be set.
30 bpm: When the auto alarm is set and the lower limit is below 30 bpm, the lower limit will be fixed to
30 bpm.
40 bpm: When the auto alarm is set and the lower limit is below 40 bpm, the lower limit will be fixed to
40 bpm.
7Set the "Buzzer Tone at Speaker Failure".
1 Press the key for "Buzzer Tone at Speaker Failure".
The dropdown list will be displayed.
2 Select from [Enable] or [Disable].
[Enable]: A buzzer tone will be generated instead of an alarm sound under the following condition.
Speaker is malfunctioning.
Alarm volume is not set to minimum.
Night mode volume is not set to [Silence].
Vital alarm (level S, H, M, L) or ventilator alarm is generating.
[Disable]: A buzzer tone will not be generated even during speaker failure.
REFERENCE
If a buzzer tone is generated at alarm generation, it can be silenced by pressing the
[Alarm Silence] key.
8Set the "Suspend Arrhy. Analysis during Noise Interference" .
1 Press the key for "Suspend Arrhy. Analysis during Noise Interference" .
The dropdown list will be displayed.
2 Select from [ON] or [OFF].
[ON]: Arrhythmia analysis will be suspended for fixed duration (5 sec.) when a noise is continuously
interfering.
[OFF]: Arrhythmia analysis will not be suspended even when a noise is continuously interfering.