5-8
Chapter 5 Initial Settings Alarm Related Setup
When "Suspend Arrhy. Analysis during Noise Interference" is set to [ON], and the
suspended duration continues for more than 30 seconds, "Cannot analyze" message
will generate.
9Set the "Lower Limit for Alarm Volume".
Changing the setting for "Alarm System" will also change the alarm volume and tone
setting. As the "Lower Limit for Alarm Volume" may also change, make sure to check the
volume and tone on the "Tone/Volume" setup menu.
1 Press the key for "Lower Limit for Alarm Volume".
The "Lower Limit for Alarm Volume" window will be displayed.
2 Set the lower limit of alarm volume for "Vital Alarm", "Ventilator Alarm", "Status Alarm".
The lower limit of adjustable alarm volume range on the "Tone/Volume" setup menu will be set. The
lower limit level can be set according to the alarm level priority, Urgent>Caution>Status.
[Test]: The test sound will be generated with the set volume.
10Set the alarm indicator operation.
The alarm indicator setting is to be performed for each alarm level.
The alarm indicator flashing pattern can be set according to the alarm level. The patient's
condition can be checked from far distance by the difference of flashing pattern.
1 Press the key for the level to set the flash pattern.
The pattern setup window will be displayed.
Alarm Indicator Flashing Pattern
Pattern Flash Pattern
Pattern A (Red, Red, Red), (xxx), (Red, Red, Red), (xxx), (Red, Red, Red)
Pattern B (Red, Orange, Red), (xxx), (Red, Orange, Red), (xxx), (Red, Orange, Red)
Pattern C (Red, Green, Red), (xxx), (Red, Green, Red), (xxx), (Red, Green, Red)
Pattern D (x , Red, x), (xxx), (x, Red, x), (xxx), (x, Red, x)
Pattern E (x, Orange, x), (xxx), (x, Orange, x), (xxx), (x, Orange, x)
Pattern F (x, Blue, x), (x, Blue, x), (x, Blue, x), (x, Blue, x), (x, Blue, x),
Pattern G (Red, Red, x), (xxx), (x, Red, Red), (xxx), (Red, Red, x)
Pattern H (Red, Orange, x), (xxx), (x, Orange, Red), (xxx), (Red, Orange, x)
Pattern I (Red, Green, x), (xxx), (x, Green, Red), (xxx), (Red, Green, x)
Pattern J (Red, x, x), (xxx), (x, Red, x), (xxx), (x, x, Red)
* (xxx) indicates that the alarm indicator is not lit.